Risk of Monopolar Electrosurgery in Cochlear Implant Recipients is Nominal: Evidence to Guide Clinical Practice
James R Dornhoffer1, Travis Haller1, Christine M Lohse2
1Department of Otolaryngology-Head and Neck Surgery, Mayo Clinic, Rochester, Minnesota, USA.
Summary
Monopolar electrosurgery is safe for cochlear implant (CI) recipients, with an extremely low risk of device complications. Electrosurgery can be used in the head and neck, avoiding direct proximity to the CI device.
Area of Science:
- Otolaryngology
- Neurosurgery
- Medical Device Safety
Background:
- Cochlear implants (CIs) are sophisticated neuroprosthetic devices.
- Electrosurgery is commonly used in head and neck procedures.
- Potential risks of electrosurgery to CI devices require assessment.
Purpose of the Study:
- To determine the prevalence of device complications from monopolar electrosurgery in cochlear implant recipients.
- To evaluate the safety of using monopolar electrosurgery in CI patients.
Main Methods:
- A multifaceted review including literature search, institutional data, manufacturer data, and surveys of high-volume centers.
- Analysis of device-cautery exposures reported in human studies and institutional records.
- Review of industry data from 3 FDA-approved CI manufacturers and surveys of 8 high-volume CI centers.
Main Results:
- Literature review identified 199 device-cautery exposures with no reported CI damage in patient records.
- Institutional data showed no CI damage in 84 instances of exposure.
- Manufacturer data revealed only one potential damage event out of 689,426 CIs; surveys found no adverse events.
Conclusions:
- The risk of adverse events or tissue injury from monopolar electrosurgery in CI recipients is extraordinarily low.
- Monopolar electrosurgery can be safely used in the head and neck of CI recipients, provided it's not in immediate proximity to the implant.
- Findings support the use of monopolar electrosurgery in CI patients and aid in patient counseling.