Related Experiment Video
Updated: Aug 17, 2026

Thermal Ablation for the Treatment of Abdominal Tumors
Published on: March 7, 2011
Feasibility and safety of bipolar radiofrequency track cautery during percutaneous image-guided abdominal biopsy
Paul B Shyn1, Adeel R Seyal2, Ravi V Gottumukkala2
1Department of Radiology, Abdominal Imaging and Intervention, Harvard Medical School, Brigham and Women's Hospital, 75 Francis St., Boston, MA, 02115, USA. pshyn@bwh.harvard.edu.
Purpose:
The purpose of this study was to assess the feasibility and safety of using a bipolar radiofrequency track cautery device during percutaneous image-guided abdominal biopsy procedures in at-risk patients.
Methods:
Forty-two patients (26-79 years old; female 44%) with at least one bleeding risk factor who underwent an abdominal image-guided (CT or US) biopsy and intended bipolar radiofrequency track cautery (BRTC) were retrospectively studied. An 18G radiofrequency electrode was inserted through a 17G biopsy introducer needle immediately following coaxial 18G core biopsy, to cauterize the biopsy track using temperature control. Bleeding risk factors, technical success, and adverse events were recorded.
Results:
BRTC was technically successful in 41/42 (98%) of procedures; in one patient, the introducer needle retracted from the liver due to respiratory motion prior to BRTC. BRTC following percutaneous biopsy was applied during 41 abdominal biopsy procedures (renal mass = 12, renal parenchyma = 10, liver mass = 9, liver parenchyma = 5, splenic mass or parenchyma = 4, gastrohepatic mass = 1). All patients had one or more of the following risk factors: high-risk organ (spleen or renal parenchyma), hypervascular mass, elevated prothrombin time, renal insufficiency, thrombocytopenia, recent anticoagulation or anticoagulation not withheld for recommended interval, cirrhosis, intraprocedural hypertension, brisk back bleeding observed from the introducer needle, or subcapsular tumor location. No severe adverse events (grade 3 or higher) occurred. Two (2/41, 5%) mild (grade 1) bleeding events did not cause symptoms or require intervention.
Conclusion:
Bipolar radiofrequency track cautery was feasible and safe during percutaneous image-guided abdominal biopsy procedures. IRB approval: MBG 2022P002277.
More Related Videos
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Ultrasonography