Orthostatic variation of pulmonary artery pressure in ambulatory heart failure patients
Prince Sethi1, Prakash Acharya1, Payton Lancaster1
1Department of Cardiovascular Disease, University of Kansas Medical Center, 3901 Rainbow Blvd, Kansas City, KS, 66160, USA.
Aim:
To study effect of change in position (supine and standing) on pulmonary artery pressure (PAP) in ambulatory heart failure (HF) patients.
Methods:
Seventeen patients with CardioMEMS® sensor and stable heart failure were consented and included in this single center study. Supine and standing measurements were obtained with at least 5 min interval between the two positions. These measurements included PAP readings utilizing the manufacturer handheld interrogator obtaining 10 s data in addition to the systemic blood pressure and heart rate recordings.
Results:
Mean supine and standing readings and their difference (Δ) were as follows respectively: Systolic PAP were 33.4 (± 11.19), 23.6 (± 10) and Δ was 9.9 mmHg (p = 0.0001), diastolic PAP were 14.2 (± 5.6), 7.9 (± 5.7) and Δ was 6.3 mmHg (p = 0.0001) and mean PAP were 21.8 (± 7.8), 14 (± 7.2) and Δ was 7.4 mmHg (p = 0.0001) while the systemic blood pressure did not vary significantly.
Conclusion:
There is orthostatic variation of PAP in ambulatory HF patients demonstrating a mean decline with standing in diastolic PAP by 6.3 mmHg, systolic PAP by 9.9 mmHg and mean PAP by 7.4 mmHg in absence of significant orthostatic variation in systemic blood pressure or heart rate. These findings have significant clinical implications and inform that PAP in each patient should always be measured in the same position. Since initial readings at the time of implant were taken in supine position, it may be best to use supine position or to obtain a baseline standing PAP reading if standing PAP is planned on being used.
Related Concept Videos
Measurement of Blood Pressure
Heart Failure II: Pathophysiology
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Heart Failure VI: Adjunct Therapies
Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement