Ex vivo biomechanical evaluation of 2.4 mm LCP plate rod constructs versus 2.7 mm LCP applied to the feline tibia
A Gutbrod1,2, F Longo2, R Affentranger3
1Tierärztliche Klinik am Hafen, Nürnberg, Germany.
Objective:
To compare the stiffness and strength of three plate and rod fixation constructs applied to a feline tibial gap model.
Study Design:
Ex vivo study.
Sample Population:
Thirty-three unpaired tibiae obtained from skeletally mature cats.
Methods:
The tibiae were randomly divided into three groups. The following implants were then applied to the feline tibiae prior to the creation of a 10 mm diaphyseal gap. Group 1: 2.4 mm locking compression plate (LCP) and 1.0 mm intramedullary pin (IMP). Group 2: 2.4 mm LCP and 1.6 mm IMP. Group 3: 2.7 mm LCP. Subsequently, each specimen was tested for torsion, axial compression, and axial load until construct failure. Student's t-tests were used to compare the torsional and axial stiffness, yield load, and maximum axial force.
Results:
Group 2 had higher axial stiffness than group 3 (p = .013). Group 1 showed a lower maximum axial force and yield point than groups 2 and 3 (p < .01; p < .05, respectively). There were no among-group differences in torsional stiffness.
Conclusion:
Constructs with a 2.4 mm LCP and 1.6 mm IMP provided the strongest and most rigid constructs in a feline tibia gap model.
Clinical Significance:
A plate-rod construct combining a 2.4 mm LCP and a 1.6 mm IMP is appropriate for achieving high implant stiffness and resisting maximum axial force in treatment of tibial fractures in cats.
More Related Videos
06:59Fracture Apparatus Design and Protocol Optimization for Closed-stabilized Fractures in Rodents
Published on: August 14, 2018
05:45Author Spotlight: Exploring Regeneration in Axolotls Through Insights in Cellular and Molecular Mechanisms, Bone Healing, and Implications for Human Therapies
Published on: April 12, 2024
Related Concept Videos
Indeterminate Structure
Applications of Stress
The...
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
Impact Loading
In cases of elastic deformation,...
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...