Automated Midline Shift and Intracranial Pressure Estimation based on Brain CT Images
Published on: April 13, 2013
Pre-acquired CT-based attenuation correction with automated headrest removal for a brain-dedicated PET system
Yuma Iwao1, Go Akamatsu2, Hideaki Tashima3
1Department of Advanced Nuclear Medicine Sciences, Institute for Quantum Medical Science, National Institutes for Quantum Science and Technology (QST), 4-9-1 Anagawa, Inage-Ku, Chiba, 263-8555, Japan. iwao.yuma@qst.go.jp.
Abstract:
Attenuation correction (AC) is essential for quantitative positron emission tomography (PET) images. Attenuation coefficient maps (μ-maps) are usually generated from computed tomography (CT) images when PET-CT combined systems are used. If CT has been performed prior to PET imaging, pre-acquired CT can be used for brain PET AC, because the human head is almost rigid. This pre-acquired CT-based AC approach is suitable for stand-alone brain-dedicated PET, such as VRAIN (ATOX Co. Ltd., Tokyo, Japan). However, the headrest of PET is different from the headrest in pre-acquired CT images, which may degrade the PET image quality. In this study, we prepared three different types of μ-maps: (1) based on the pre-acquired CT, where namely the headrest is different from the PET system (μ-map-diffHr); (2) manually removing the headrest from the pre-acquired CT (μ-map-noHr); and (3) artificially replacing the headrest region with the headrest of the PET system (μ-map-sameHr). Phantom images by VRAIN using each μ-map were investigated for uniformity, noise, and quantitative accuracy. Consequently, only the uniformity of the images using μ-map-diffHr was out of the acceptance criteria. We then proposed an automated method for removing the headrest from pre-acquired CT images. In comparisons of standardized uptake values in nine major brain regions from the 18F-fluoro-2-deoxy-D-glucose-PET of 10 healthy volunteers, no significant differences were found between the μ-map-noHr and the μ-map-sameHr. In conclusion, pre-acquired CT-based AC with automated headrest removal is useful for brain-dedicated PET such as VRAIN.
More Related Videos
05:32A High-performance Compact Photoacoustic Tomography System for In Vivo Small-animal Brain Imaging
Published on: June 21, 2017
09:07Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic Stimulation (TMS).