Safety and function of programmable ventriculo-peritoneal shunt valves: An in vitro 7 Tesla magnetic resonance
Bixia Chen1,2, Philipp Dammann1,2, Ramazan Jabbarli2
1Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Essen, North Rhine Westphalia, Germany.
Objective:
The quantity of ultra-high field MRI neuroimaging studies has rapidly increased. This study tests function, safety, and image artifacts of two frequently implanted programmable ventriculo-peritoneal (VP) shunt valves in a 7T MRI system.
Methods:
All tests were performed using a whole-body 7T MRI system. Three proGAV 2.0 and 3 CODMAN CERTAS® Plus programmable VP-shunt valves were tested in three steps. 1) Deflection angle tests close to the bore opening at the location of a static magnetic field gradient of 3-5 T/m. 2) Valves were fixed on a spherical phantom in 3 positions (a. lateral, b. cranial, c. cranial with 22.5° tilt anteriorly) and assessed for keeping the programmed pressure setting and reprogrammability. 3) Valves were fixed on the phantom and positioned lateral in a radiofrequency head coil. MRI scans were performed for both models, including MPRAGE, GRE and SE sequences.
Results:
Deflection angles were moderate (13°, 14°, 13°) for the proGAV valves and close to critical (43°, 43°, 41°) for the CODMAN valves at the test location. Taking a scaling factor of 2-3 for the maximum spatial magnetic field gradient accessible to a patient within the magnet bore into account renders both valves MR unsafe regarding ferromagnetic attraction. The proGAV valves kept the pressure settings in all positions and were reprogrammable in positions a. and b. In position c., reprogrammability was lost. The CODMAN valves changed their pressure setting and reprogrammability was lost in all positions. MR image signal homogeneity was unaltered in the phantom center, artifacts limit the assessability of structures in close vicinity to the valves.
Conclusion:
Both tested programmable VP-shunt valves are MR unsafe for 7T systems. Novel programming mechanisms using permanent magnets with sufficient magnetic coercivity or magnet-free mechanisms may allow the development of programmable VP-shunt valves that are conditional for 7T MR systems.
Related Concept Videos
Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging