Test and Development of a Specialized Pipeline for Ventilator Calibration
Junming Liu1, Wenxiu Zheng2, Ziyan Wang3
1Tai'an Institute of Quality and Technology Inspection and Testing, Taian, 271000, China.
Abstract:
A specialized Y-type pipeline is developed to replace medical disposable Y-type pipelines for ventilator calibration and to improve accuracy, providing a reference for improvement. According to the calibration specifications of ventilators, tests are performed on specialized pipelines and more to compare their data. Taking the tidal volume 400-mL detection point as an example, the calibration errors made by medical disposable pipelines and specialized pipelines are 6.2% and -0.8%, respectively, and the errors at other detection points are roughly the same. The accuracy of the specialized Y-type pipeline has significantly improved, reducing the impact of the pipeline on tidal volume calibration from more than 6% to less than 1%. The use of specialized pipelines can significantly improve the accuracy and qualification rate of calibration, reducing the consumption of disposable Y-type pipelines and thereby significantly reducing costs and increasing efficiency.
More Related Videos
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...