Audit of data from examination image headers collected for quality assurance in the ECOG-ACRIN EA1151 tomosynthesis
Aili K Maki1, Gordon E Mawdsley1, James G Mainprize1
1Physical Sciences, Sunnybrook Research Institute, Toronto, Ontario, Canada.
Purpose:
A comprehensive, centrally-monitored physics quality control (QC) program was developed for the Tomosynthesis Imaging Screening Trial (TMIST), a randomized controlled trial of digital breast tomosynthesis (TM) versus digital mammography (DM) for cancer screening. As part of the program, in addition to a set of phantom-based tests, de-identified data on image acquisition and processing parameters were captured from the DICOM headers of all individual patient images in the trial. These data were analyzed to assess the potential usefulness of header data from digital mammograms and tomosynthesis images of patients for quality assurance in breast imaging.
Methods:
Data were automatically extracted from the headers of all de-identified patient mammograms and tomosynthesis images in the TMIST study. Image acquisition parameters and estimated radiation doses were tracked for individual sites, systems and across system types. These parameters included (among others) kV, target/filter use, number of acquired views per examination, AEC mode, compression thickness and force and detector temperature. Consistency of manually entered study data parameters (subject ID, screening time-point) from TMIST was evaluated. Preliminary observations from the program are presented.
Results:
We report on data from 812 651 images from 135 525 examinations acquired between October, 2017 and December, 2022. Data came from 6 system models from 3 manufacturers. There was greater variability both in the number of views used and in the estimated (proxy) doses received in DM exams compared to TM. Mean proxy doses per examination varied among manufacturers from 2.76-4.54 mGy for DM and 3-4.84 mGy for the tomosynthesis component in the TM arm with maximum examination proxy doses of 20 and 26 mGy for DM and TM respectively. Mean proxy doses per examination for the combination examination in TM (tomosynthesis plus digital mammography) varied from 6.6 to 7.6 mGy among manufacturers with a maximum of 44.5 mGy.
Conclusions:
Overall, modern digital mammography and tomosynthesis systems used in TMIST have operated very reliably. Doses vary considerably due to variation in the number of views per examination, thickness and fibro-glandularity of the breast, and choices in the use of synthesized versus actual 2D mammography in the TM examination. These data may also be useful in predicting equipment problems. Header information is valuable not only for automated QC, but also for cross-checking accuracy and consistency of data in a clinical study.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
13:07Optical Coherence Tomography Based Biomechanical Fluid-Structure Interaction Analysis of Coronary Atherosclerosis Progression
Published on: January 15, 2022
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT