Effects of Synchronizing Foot Strike and Cardiac Phase on Exercise Hemodynamics in Patients With Cardiac
Denis J Wakeham1,2, Erika Ivey2, Sophie A Saland2
1University of Texas Southwestern Medical Center, Dallas (D.J.W., T.L.B., C.M.H., S.S., J.P.M., M.H., B.D.L.).
Background:
Despite advances in medical and cardiac resynchronization therapy (CRT), individuals with chronic congestive heart failure (CHF) have persistent symptoms, including exercise intolerance. Optimizing cardio-locomotor coupling may increase stroke volume and skeletal muscle perfusion as previously shown in healthy runners. Therefore, we tested the hypothesis that exercise stroke volume and cardiac output would be higher during fixed-paced walking when steps were synchronized with the diastolic compared with systolic portion of the cardiac cycle in patients with CHF and CRT.
Methods:
Ten participants (58±17 years of age; 40% female) with CHF and previously implanted CRT pacemakers completed 5-minute bouts of walking on a treadmill (range, 1.5-3 mph). Participants were randomly assigned to first walking to an auditory tone to synchronize their foot strike to either the systolic (0% or 100±15% of the R-R interval) or diastolic phase (45±15% of the R-R interval) of their cardiac cycle and underwent assessments of oxygen uptake (V̇o2; indirect calorimetry) and cardiac output (acetylene rebreathing). Data were compared through paired-samples t tests.
Results:
V̇o2 was similar between conditions (diastolic 1.02±0.44 versus systolic 1.05±0.42 L/min; P=0.299). Compared with systolic walking, stroke volume (diastolic 80±28 versus systolic 74±26 mL; P=0.003) and cardiac output (8.3±3.5 versus 7.9±3.4 L/min; P=0.004) were higher during diastolic walking; heart rate (paced) was not different between conditions. Mean arterial pressure was significantly lower during diastolic walking (85±12 versus 98±20 mm Hg; P=0.007).
Conclusions:
In patients with CHF who have received CRT, diastolic stepping increases stroke volume and oxygen delivery and decreases afterload. We speculate that, if added to pacemakers, this cardio-locomotor coupling technology may maximize CRT efficiency and increase exercise participation and quality of life in patients with CHF.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
Related Concept Videos
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Electrophysiology of Normal Cardiac Rhythm
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Cardiac Cycle
During the cardiac cycle, blood flow through the heart is regulated entirely by changing pressure gradients. This sequence of events begins with the heart in a state of total relaxation, known as mid-to-late diastole, during which blood passively flows from...
Heart Failure II: Pathophysiology