Related Experiment Video
Updated: Jul 13, 2025

TAPE: A Biodegradable Hemostatic Glue Inspired by a Ubiquitous Compound in Plants for Surgical Application
Published on: June 8, 2016
A temperature-sensitive, high-adhesion medical tape: a comparative, single-blind clinical trial
Shawn Swanson1, Vivian Luu1, Ryan Smith1
1Human Photonics Laboratory, Department of Mechanical Engineering, University of Washington, Seattle, WA, US.
Objective:
Medical adhesives are used to secure wound care dressings and other critical devices to the skin. While high peel-strength adhesives provide more secure skin attachment, they are difficult to remove from the skin and are correlated with medical adhesive-related skin injuries (MARSI), including skin tears, and an increased risk of infection. Lower-adhesion medical tapes may be applied to avoid MARSI, leading to dressing or device dislodgement and further medical complications.
Method:
This paper reports on the clinical testing of a new, high-adhesion medical tape, ThermoTape (University of Washington, US), designed for low skin trauma upon release. ThermoTape was benchmarked with Tegaderm (3M, US) and Kind Removal Tape (KRT) (3M, US). All three tapes were applied to both the left and right forearm of healthy volunteers and were removed 24 hours later-the right arm without applying heat and the left arm by applying a heat pack for 30 seconds before removal. Tape wear, self-reported pain (0-10 scale) and skin redness 15 minutes after removal were recorded.
Results:
This was a 53-subject comparative, single-blind clinical trial. There were clinically and statistically significant results supporting reduced pain during removal of ThermoTape with warming, with an average 58% decrease in pain, paired with a statistically significant 45% reduction in skin redness (p<0.01 for both values). In contrast, there were statistically insignificant differences in pain and redness for removal of Tegaderm and KRT with warming. ThermoTape after warming, in comparison with Tegaderm without warming, produced a reduced pain score of >1 on the 0-10 Wong-Baker/Face pain scale, which was statistically significant (p<0.01).
Conclusion:
These results provide compelling evidence that warming ThermoTape prior to removal can reduce pain and injury when compared with standard medical tapes. This could allow for stronger attachment of wound care dressings and critical medical devices while reducing cases of MARSI.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...