Related Experiment Video
Updated: Jul 13, 2025

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Impact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients
Dominik Lisowski1, Paul Lutyj1, Arya Abazari1
1Department of Radiation Oncology, University Hospital Würzburg, 97080 Würzburg, Germany.
Purpose:
This study analyses a large number of cancer patients with CIEDs for device malfunction and premature battery depletion by device interrogation after each radiotherapy fraction and compares different guidelines in regard to patient safety.
Methods:
From 2007 to 2022, a cohort of 255 patients was analyzed for CIED malfunctions via immediate device interrogation after every RT fraction.
Results:
Out of 324 series of radiotherapy treatments, with a total number of 5742 CIED interrogations, nine device malfunctions (2.8%) occurred. Switching into back-up/safety mode and software errors occurred four times each. Once, automatic read-out could not be performed. The median prescribed cumulative dose at planning target volume (PTV) associated with CIED malfunction was 45.0 Gy (IQR 36.0-64.0 Gy), with a median dose per fraction of 2.31 Gy (IQR 2.0-3.0 Gy). The median maximum dose at the CIED at time of malfunction was 0.3 Gy (IQR 0.0-1.3 Gy). No correlation between CIED malfunction and maximum photon energy (p = 0.07), maximum dose at the CIED (p = 0.59) nor treatment localization (p = 0.41) could be detected. After excluding the nine malfunctions, premature battery depletion was only observed three times (1.2%). Depending on the national guidelines, 1-9 CIED malfunctions in this study would have been detected on the day of occurrence and in none of the cases would patient safety have been compromised.
Conclusion:
Radiation-induced malfunctions of CIEDs and premature battery depletion are rare. If recommendations of national safety guidelines are followed, only a portion of the malfunctions would be detected directly after occurrence. Nevertheless, patient safety would not be compromised.