Robust voltage-controlled transcutaneous energy transfer system for artificial anal sphincter
Yelin Chen1,2, Pingping Jiang1,2, Lichao Wang1,2
1Shanghai Engineering Research Center of Intelligent Addiction Treatment and Rehabilitation, Shanghai Jiao Tong University, Shanghai, China.
Background:
The artificial anal sphincter (AAS) system has gained significant attention as a solution for treating fecal incontinence (FI). It relies on transcutaneous energy transfer (TET) as its primary energy source. However, changes in posture or biological tissue can cause misalignment of the coil, resulting in unstable power reception. Inadequate power affects charging efficiency, while excessive power leads to excessive heating at the receiver side. Consequently, achieving safe and constant voltage charging for the AAS becomes a complex challenge.
Methods:
To maintain a consistent charging voltage and overcome the issue of variations in load and coil coupling strength, this article proposes a wireless charging control system that utilizes an LCC-S-type resonant network and phase shift to adjust the transmitting voltage based on feedback charging voltage in real time. In particular, the PI controller and neural network are introduced to change the phase-shift angle swiftly. The dynamic performance is then evaluated under different misalignments and presented with comparative results.
Results:
The results indicate that the multilayer perceptron control system outperforms the PI. Under the complex misalignment disturbance, the average error of receiver side load voltage is only 0.007 V, with an average settling time of 960 ms. Additionally, the average temperature at the receiver side is 40.4°C.
Conclusion:
The experiments demonstrate that the proposed system effectively addresses the misalignment issue in TET during the charging, ensuring constant voltage charging at the receiver side and thermal safety.
Related Concept Videos
Electrical Synapses
Gap junctions allow the current to pass directly from one cell to the next. In contrast, in the chemical synapse, the neurotransmitters carry the information through the synaptic cleft from one neuron to the next. They consist of two...
Sensory Functions of the Skin
There are two main categories of receptors on the skin: capsulated and non-capsulated. The non-capsulated ones are mainly the pain receptors. The capsulated ones can be further categorized based on the...
Tension
Muscle Stimulation Frequency
Wave summation
At low firing rates, motor neurons induce individual twitch contractions in muscle fibers. These twitches...
Functions of the Nervous System
Nervous System
Extending...