A new interpretive method for impedance plethysmography.

W R Lynch

    Medical Instrumentation
    |September 1, 1986
    PubMed
    Summary
    This summary is machine-generated.

    A new interpretative method for impedance plethysmography significantly improves deep venous thrombosis diagnosis by reducing false negatives and borderline results compared to Wheeler's method.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Effect of rate and inspiratory flow on ventilator-induced lung injury.

    The Journal of trauma·2000
    Same author

    Hemodynamic effect of a low-resistance artificial lung in series with the native lungs of sheep.

    The Annals of thoracic surgery·2000
    Same author

    Indiana's Emergency Medical Services. An open letter to physicians interested in improving pre-hospital (ambulance) patient care.

    Indiana medicine : the journal of the Indiana State Medical Association·1985
    Same author

    Hydroxyproline metabolism in two sisters with hydroxyprolinemia.

    Human heredity·1979
    Same journal

    Nurses' attitudes toward conventional and automated vital signs measurement methods.

    Medical instrumentation·1988
    Same journal

    Implementation of a home-based program for early detection of clinical deterioration in cystic fibrosis.

    Medical instrumentation·1988
    Same journal

    A home-based pulmonary function monitor for cystic fibrosis.

    Medical instrumentation·1988
    Same journal

    Blood pressure reactivity and bias vary with age in a comparison of traditional and automated methods of measurement.

    Medical instrumentation·1988
    Same journal

    Human factors and design evaluation of digital blood pressure/pulse meters.

    Medical instrumentation·1988
    Same journal

    Electrically produced artificial ventilation.

    Medical instrumentation·1988
    See all related articles

    Area of Science:

    • Vascular Medicine
    • Diagnostic Imaging
    • Medical Technology

    Background:

    • Wheeler's interpretative method for impedance plethysmography often yields false negative and borderline results.
    • Deep venous thrombosis (DVT) diagnosis requires accurate and reliable interpretation of imaging studies.

    Purpose of the Study:

    • To compare the diagnostic accuracy of a new impedance plethysmography (IP) interpretative method against Wheeler's method.
    • To evaluate the efficacy of the new method in reducing false negative and borderline interpretations in suspected DVT cases.

    Main Methods:

    • A comparative study involving 21 patients with suspected DVT.
    • Impedance plethysmography (IP) was performed, and interpretations from Wheeler's method and a new method were compared with venography results.

    Related Experiment Videos

  • Patients were categorized based on venogram findings: no thrombosis, tibial venous thrombosis, or proximal DVT.
  • Main Results:

    • The new method achieved 75% true-positive interpretations for DVT, with no false negatives, compared to Wheeler's method's 8% true-positive and 17% false-negative rates.
    • The new method reduced borderline interpretations from 62% to 33%.
    • Both methods showed 56% true-negative interpretations for patients without DVT.

    Conclusions:

    • The novel impedance plethysmography interpretative method enhances DVT diagnosis by significantly lowering borderline and false-negative interpretations.
    • The new method provides valuable guidelines for identifying patients with borderline results who may require further diagnostic testing.