Optimal interlesion distance for 90 and 50 watt radiofrequency applications with low ablation index values:
Agustín Alfonso Bortone1, F Daniel Ramirez2,3, Marion Constantin4,5
1ELSAN, Service de Cardiologie, Hôpital Privé Les Franciscaines, 3 Rue Jean Bouin, 30000 Nîmes, France.
Aims:
The optimal interlesion distance (ILD) for 90 and 50 W radiofrequency applications with low ablation index (AI) values in the atria has not been established. Excessive ILDs can predispose to interlesion gaps, whereas restrictive ILDs can predispose to procedural complications. The present study sought, therefore, to experimentally determine the optimal ILD for 90 W-4 s and 50 W applications with low AI values to optimize catheter ablation outcomes in humans.
Methods And Results:
Posterior intercaval lines were created in eight adult sheep using CARTO and the QDOT-MICRO catheter in a temperature-controlled mode. In four animals, the lines were created with 50 W applications, a target AI value ≥350, and ILDs of 6, 5, 4, and 3 mm, respectively. In the other four animals, the lines were created with 90 W-4 s applications and ILDs of 6, 5, 4, and 3 mm, respectively. Activation maps were created immediately after ablation and at 21 days to assess linear block prior to gross and histological analyses. All eight lines appeared transmural and continuous on histology. However, for 50 W-only applications with an ILD of 3 mm resulted in durable linear electrical block, whereas for 90 W applications, only the lines with ILDs of 4 and 3 mm were blocked. No complications were detected during ablation procedures, but all power and ILD combinations except 50 W-6 mm resulted in asymptomatic shallow lung lesions.
Conclusion:
In the intercaval region in sheep, for 50 W applications with an AI value of ∼370, the optimal ILD is 3 mm, whereas for 90 W-4 s applications, the optimal ILD is 3-4 mm.
More Related Videos
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
Related Concept Videos
Generating Electromagnetic Radiations
Electromagnetic Waves in Matter
Consider the electromagnetic wave passing through a dielectric medium. In such a case, Maxwell's equations get modified. In Ampere's law, ε0 , the dielectric permittivity of free space is replaced with ε, the permittivity of dielectric. Also, the vacuum permeability μ0 is replaced by the permeability of the medium, μ.
Furthermore, the...
Standing Waves in a Cavity
Characteristics of Series Resonant Circuit
Lossless Lines
Boundary Conditions: Lossless Lines
At the receiving end, the boundary condition states that the voltage equals the product of the receiving-end impedance and current. This relationship is expressed as a function of the incident and...