Related Experiment Video
Updated: Jul 12, 2025

Computed Tomography-guided Time-domain Diffuse Fluorescence Tomography in Small Animals for Localization of Cancer Biomarkers
Published on: July 17, 2012
Photoacoustic tomography as a method to estimate the optical fluence distribution in turbid media
Lawrence C M Yip1,2, Elina Rascevska1,3, Parsa Omidi3
1Lawson Health Research Institute, Imaging Program, 268 Grosvenor St., London, Canada, N6A 4V2, Canada.
Abstract:
Currently, there are no non-invasive experimental methods available for measuring optical fluence distributions in tissue. We present photoacoustic tomography (PAT) as a method to approximate the relative optical fluence distribution in a homogeneous optically scattering medium. Three-dimensional photoacoustic images were captured with a near-full view PAT scanner in phantoms with known optical absorption and scatter properties. Resultant 3D PAT images were compared to the expected optical fluence distributions from Monte Carlo simulations and diffusion theory using volumetric and shape analysis. Volumetric analysis of PAT images compared well with the optical fluence distributions from simulation. Dice similarity coefficients ranged from 51 to 82%. The reduced scattering coefficient estimated from PAT images compared well to estimates from simulations for values below 0.5 mm-1. Near full-view PAT has been found to be useful for estimating the optical fluence distribution in an optically scattering medium. Further development is needed to extend the measurement range.