Related Experiment Video
Updated: Jul 12, 2025

Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
Accelerated Cine Cardiac MRI Using Deep Learning-Based Reconstruction: A Systematic Evaluation
Amol Pednekar1,2, Murat Kocaoglu1,2, Hui Wang1,2,3
1Department of Radiology, Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio, USA.
Background:
Breath-holding (BH) for cine balanced steady state free precession (bSSFP) imaging is challenging for patients with impaired BH capacity. Deep learning-based reconstruction (DLR) of undersampled k-space promises to shorten BHs while preserving image quality and accuracy of ventricular assessment.
Purpose:
To perform a systematic evaluation of DLR of cine bSSFP images from undersampled k-space over a range of acceleration factors.
Study Type:
Retrospective.
Subjects:
Fifteen pectus excavatum patients (mean age 16.8 ± 5.4 years, 20% female) with normal cardiac anatomy and function and 12-second BH capability.
Field Strength/Sequence:
1.5-T, cine bSSFP.
Assessment:
Retrospective DLR was conducted by applying compressed sensitivity encoding (C-SENSE) acceleration to systematically undersample fully sampled k-space cine bSSFP acquisition data over an acceleration/undersampling factor (R) considering a range of 2 to 8. Quality imperceptibility (QI) measures, including structural similarity index measure, were calculated using images reconstructed from fully sampled k-space as a reference. Image quality, including contrast and edge definition, was evaluated for diagnostic adequacy by three readers with varying levels of experience in cardiac MRI (>4 years, >18 years, and 1 year). Automated DL-based biventricular segmentation was performed commercially available software by cardiac radiologists with more than 4 years of experience.
Statistical Tests:
Tukey box plots, linear mixed effects model, analysis of variance (ANOVA), weighted kappa, Kruskal-Wallis test, and Wilcoxon signed-rank test were employed as appropriate. A P-value <0.05 was considered statistically significant.
Results:
There was a significant decrease in the QI values and edge definition scores as R increased. Diagnostically adequate image quality was observed up to R = 5. The effect of R on all biventricular volumetric indices was non-significant (P = 0.447).
Data Conclusion:
The biventricular volumetric indices obtained from the reconstruction of fully sampled cine bSSFP acquisitions and DLR of the same k-space data undersampled by C-SENSE up to R = 5 may be comparable.
Evidence Level:
3 TECHNICAL EFFICACY: Stage 1.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT