Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Non-radioactive electron capture detector based on soft X-ray ionization for gas chromatography - A possible
Maximilian J Kueddelsmann1, Jannie J Stadtler1, Ansgar T Kirk1
1Institute of Electrical Engineering and Measurement Technology, Department of Sensors and Measurement Technology, Leibniz University Hannover, Appelstr. 9A, Hannover 30167, Germany.
Abstract:
In this paper, we present a new electron capture detector based on a compact X-ray tube (X-ECD) for electron generation by soft X-ray radiation instead of using a radioactive source. ECDs are commonly used in many laboratories as standard GC detectors since their invention in the 1950s, especially for highly sensitive detection of halogenated substances, pesticides or other environmental pollutants. However, due to unsatisfactory alternatives, many ECDs are still used with radioactive β-emitters, which is difficult and expensive in most applications today due to legal restrictions. The new X-ECD contains a small X-ray tube for generating free electrons by ionizing the carrier gas like in radioactive ECDs. Thus, no additional dopants or special gases are required. The X-ECD has limits of detection in the pptv range and shows linearity over a wide concentration range. Furthermore, the used X-ray tube shows good long-term stability. So far, we have operated the X-ray tube continuously for about one year without notable degradation. However, in case of future degradation, the X-ECD can still be operated with the same sensitivity by simple adjusting the set point current in constant current mode. This makes calibration robust against possible degradation of the X-ray tube. In combination with a conventional gas chromatograph, the X-ECD is able to detect halogenated hydrocarbons and even low volatile pesticides without any peak distortion such as tailing. Thereby a minimum detectability in the upper fg/µL range for Lindane was reached, which is similar when compared to radioactive ECDs.
More Related Videos
07:55Elemental-sensitive Detection of the Chemistry in Batteries through Soft X-ray Absorption Spectroscopy and Resonant Inelastic X-ray Scattering
Published on: April 17, 2018
06:46Applying X-ray Imaging Crystal Spectroscopy for Use as a High Temperature Plasma Diagnostic
Published on: August 25, 2016
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...