The image quality and diagnostic performance of CT perfusion-derived CT angiography versus that of conventional CT
Xiu-Zhi Zhou1, Kuan Lu1, Du-Chang Zhai1
1Department of Radiology, The Second Affiliated Hospital of Soochow University, Suzhou, China.
Background:
The combination of computed tomography angiography (CTA) and computed tomography perfusion (CTP) evaluation of cerebral perfusion status and vascular conditions can improve the diagnostic accuracy of infarction, ischemia, and vascular occlusion in stroke patients, as well as a comprehensive assessment of cerebral edema, collateral circulation, and blood perfusion in the lesion area. However, the consequent radiation safety and contrast agent nephropathy have aroused increasing concern. The purpose of this study was to assess the image quality and diagnostic accuracy of CTA images derived from CTP data, and to explore the feasibility of replacing conventional CTA.
Methods:
A total of 31 consecutive patients with suspected acute ischemic stroke were retrospectively analyzed. All patients underwent head and neck CTA and brain CTP examinations. All the CTP images were transmitted to the ShuKun artificial intelligence system, which reconstructs CTA derived from CTP (CTA-DF-CTP). The images were divided into 2 groups, including CTA-DF-CTP (Group A) and conventional CTA (Group B). The CT attenuation values, subjective image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), image quality, CT volume dose index (CTDIvol), dose length product (DLP), and effective radiation dose (ED) were compared between the 2 groups. Moreover, the consistency of vascular stenosis and stenosis degree between the 2 groups were measured and evaluated.
Results:
There were no significant differences in image noise, SNR, or CNR between Groups A and B (P>0.05). The CT attenuation values of the arteries were higher in Group A than in B [internal carotid artery (ICA) =548±112 vs. 454±85 Hounsfield units (HU), middle cerebral artery (MCA) =453±118 vs. 388±70 HU, and basilar artery (BA) =431±99 vs. 360±83 HU] (P<0.01). The image quality of the 2 groups met the requirement of clinical diagnosis (4.97±0.18 vs. 4.94±0.25). No significant difference was found in subjective evaluation (P>0.05). In Group A compared with Group B, the following reductions were observed: CTDIvol (10.7%; 100.8 vs. 112.9 mGy), DLP (23.0%; 1,613±0 vs. 2,093±88 mGy·cm), and ED (23.0%; 5.00±0.00 vs. 6.49±0.27 mSv).
Conclusions:
CTA-DF-CTP data provide diagnostic accuracy and image quality similar to those of conventional CTA of head and neck CTA.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies VII: Vascular Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...