Assessing the manufacturable 32-channel cochlear electrode array: evaluation results for clinical trials.
Gwangjin Choi1, Yoonhee Ha1, Doo-Hee Kim1
1TODOC Co., Ltd., Seoul, 08394, South Korea.
Biomedical Microdevices
|October 23, 2023
Summary
This study reports on a new 32-channel cochlear electrode array. The device passed mechanical, electrical, and histological reliability tests, showing promise for cochlear implant systems.
Area of Science:
- Biomedical Engineering
- Materials Science
- Otolaryngology
Background:
- Cochlear implants require reliable electrode arrays for effective hearing restoration.
- Existing manufacturing methods may limit scalability and precision.
- Development of advanced electrode arrays is crucial for improving cochlear implant performance.
Purpose of the Study:
- To report reliability evaluation results of a newly manufactured 32-channel cochlear electrode array.
- To verify the manufacturability and performance of the electrode array according to international standards.
- To assess the insertion trauma associated with the electrode array in human temporal bone studies.
Main Methods:
- Automated laser micro-machining and layer-by-layer silicone deposition for electrode array fabrication.
- Testing according to ISO Standard 14708-7, including elongation, insulation, flexural, and fatigue tests.
- Human temporal bone studies involving insertion via the round window and histological assessment.
Main Results:
- The electrode array met ISO 14708-7 requirements, maintaining electrical continuity and insulation post-testing.
- Impedance measurements remained above the 100 kΩ pass/fail criterion after saline soaking.
- Human temporal bone studies showed an atraumatic insertion rate of 60.3% (grade 0 or 1) with an average insertion angle of 399.7°.
Conclusions:
- The developed 32-channel cochlear electrode array demonstrates verified reliability in mechanical, electrical, and histological aspects.
- The manufacturing methods enable fine patterning and mass production, suitable for clinical application.
- Further evaluation in clinical trials is planned to assess the performance and stability of the 32-channel cochlear implant system.