Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Lower Revision Rates and Improved Stability With a Monoblock Ceramic Acetabular Cup
Loïc Y G Vercruysse1, Lachlan P Milne2, Dylan T C Harries3
1Department of Orthopaedic Surgery, Sir Charles Gairdner Hospital, Nedlands, Western Australia, Australia; Department of Development and Regeneration KU Leuven, Institute for Orthopaedic Research and Training KU Leuven, Leuven, Belgium.
The DeltaMotion ceramic cup showed a low revision rate for total hip arthroplasty, including dislocation and ceramic breakage. This monoblock design offers potential benefits for active patients, though it is no longer manufactured.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Arthroplasty Research
Background:
- Monoblock ceramic cups with large-diameter femoral heads aim to improve range of motion and stability in total hip arthroplasty (THA).
- These features are particularly beneficial for younger, high-demand patients requiring THA.
- The DeltaMotion cup, a specific monoblock ceramic design, was evaluated for its performance.
Purpose of the Study:
- To assess the survival rate of the DeltaMotion cup.
- To identify the primary reasons for revision surgery in patients who received the DeltaMotion cup.
- To compare the performance of the DeltaMotion cup against modular bearing designs.
Main Methods:
- Analysis of Australian National Joint Replacement Registry (AOANJRR) data for primary conventional THA from 2001-2021.
- Inclusion of prostheses with ceramic/ceramic, ceramic/XLPE, metal/XLPE, or CM/XLPE bearing surfaces.
- Primary outcome: cumulative percent revision for all causes; Secondary outcomes: revision for dislocation, ceramic breakage, or noise.
Main Results:
- Out of 486,946 primary THA procedures, 4,033 used the DeltaMotion cup, and 482,913 used modular designs.
- The DeltaMotion cup demonstrated the lowest cumulative percent revision for all diagnoses compared to modular bearings.
- Significantly lower revision rates for dislocation and no revisions for squeaking were observed with the DeltaMotion cup; only 1 ceramic breakage occurred versus 175 in modular groups.
Conclusions:
- The DeltaMotion cup exhibited a low rate of all-cause revision, dislocation, ceramic breakage, and noise.
- While no longer manufactured, its performance provides insights for future monoblock ceramic cup development.
- Further follow-up of contemporary monoblock ceramic cups is necessary to confirm their suitability for active patient populations.