HYPER: pre-clinical device for spatially-confined magnetic particle hyperthermia
Hayden Carlton1, Matthias Weber2, Maximilian Peters2
1Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD, USA.
Purpose:
Magnetic particle hyperthermia is an approved cancer treatment that harnesses thermal energy generated by magnetic nanoparticles when they are exposed to an alternating magnetic field (AMF). Thermal stress is either directly cytotoxic or increases the susceptibility of cancer cells to standard therapies, such as radiation. As with other thermal therapies, the challenge with nanoparticle hyperthermia is controlling energy delivery. Here, we describe the design and implementation of a prototype pre-clinical device, called HYPER, that achieves spatially confined nanoparticle heating within a user-selected volume and location.
Design:
Spatial control of nanoparticle heating was achieved by placing an AMF generating coil (340 kHz, 0-15 mT), between two opposing permanent magnets. The relative positions between the magnets determined the magnetic field gradient (0.7 T/m-2.3 T/m), which in turn governed the volume of the field free region (FFR) between them (0.8-35 cm3). Both the gradient value and position of the FFR within the AMF ([-14, 14]x, [-18, 18]y, [-30, 30]z) mm are values selected by the user via the graphical user interface (GUI). The software then controls linear actuators that move the static magnets to adjust the position of the FFR in 3D space based on user input. Within the FFR, the nanoparticles generate hysteresis heating; however, outside the FFR where the static field is non-negligible, the nanoparticles are unable to generate hysteresis loss power.
Verification:
We verified the performance of the HYPER to design specifications by independently heating two nanoparticle-rich areas of a phantom placed within the volume occupied by the AMF heating coil.
More Related Videos
06:45In Vitro and In Vivo Delivery of Magnetic Nanoparticle Hyperthermia Using a Custom-Built Delivery System
Published on: July 2, 2020
13:41Magnetic Resonance-Guided High Intensity Focused Ultrasound Generated Hyperthermia: A Feasible Treatment Method in a Murine Rhabdomyosarcoma Model
Published on: January 13, 2023
Related Concept Videos
Decreased Body Temperature
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...