Electromagnetically Navigated Tube Placement Device for Bedside Placement of Small Bowel Feeding Tube on Patients
Guoliang Tan1, Yongming Chen1, Yanping Lin1
1Department of Critical Care Medicine, The Second Affiliated Hospital of Fujian Medical University, No. 34, Zhongshan North Road, Licheng District, Quanzhou City, Fujian, 362000, China.
Background:
A developing approach for the bedside installation of feeding tubes is the Electromagnetic Navigation-assisted Tube Placement Device (ENTPD). The ENTPD monitors the tip position of feeding tubes when they are inserted into the digestive tract. It aids in the avoidance of airway misalignment and allows placing into the small bowel. Several recent exploratory studies have shown that ENTPD for nasojejunal feeding tube installation can improve success rates, lower costs, and allow for a more rapid beginning of enteral nutrition.
Objectives:
The aim of this study was to compare the effect of using an ENTPD for bedside placement of small bowel feeding tubes with blind placement on patients with acute severe pancreatitis and to see how well the electromagnetic navigation trajectory image (ENTI) and X-ray agreed on the location of the tube tip after placement.
Methods:
The study was done prospectively using randomized and single-blind methods. The 65 cases used electromagnetic navigation-assisted placement, and 58 cases were blind placement. For judging the tube tip location, we compared the success rate, median time, number of repeat placements, complications, and agreement of ENTI vs. X-ray.
Results:
The blind placement group's success rate was 86.21% compared to the ENTPD's 95.38%, P = 0.075. The median time was significantly longer in the blind placement group (116.55 ± 68.62 min vs. 25.37 ±12.63 min, P=0.000); the average number of repeating placements was 3.02 ± 1.21 vs. 1.16 ± 0.31 (Blind placement vs. ENTPD, P = 0.002). It provided a high degree of agreement between ENTI and X-ray after contrast, κ=0.752 [95% confidence interval, 0.67-0.84]. No complications occurred in the two groups.
Conclusion:
ENTPD was used safely and effectively at the bedside to help patients with acute severe pancreatitis get feeding tubes. It not only improved the high successful rate of placement, decreased the time and reduced the exposure to X-ray, but it was also very convenient for bedside placement because of the portable equipment.
Related Concept Videos
Enteral Nutrition I: Orogastric and Nasogastric Feeding
Orogastric (OG) and nasogastric (NG) feeding are two standard methods used for enteral nutrition. Enteral nutrition is often preferred over...
Enteral Nutrition II: Nasointestinal and Gastrostomy Feeding
Nasointestinal Feeding
Nasointestinal feeding involves placing a tube...
Endoscopic Procedures V: ERCP
Patient...