Related Experiment Video
Updated: Jul 12, 2025

06:17
Augmented Reality Navigation-Guided Core Decompression for Osteonecrosis of Femoral Head
Published on: April 12, 2022
3.7K
Poor usability of computer-assisted navigation for hip fracture surgery
Rasmus Abildtrup Hestehave1, Per Hviid Gundtoft1, Christian Lind Nielsen1
1Department of Orthopaedics, Aarhus University Hospital, Palle Juul-Jensens Boulevard 99, J801, 8200, Aarhus, Denmark.
Archives of Orthopaedic and Trauma Surgery
|October 25, 2023
Summary
The Stryker ADAPT system for Gamma3 nail surgery received "poor" usability ratings from surgeons. Its use was discontinued at the institution due to these findings.
Area of Science:
- Orthopaedic Surgery
- Medical Device Technology
- Surgical Navigation Systems
Background:
- The Stryker ADAPT computer-assisted navigation system offers intraoperative feedback for Gamma3 nail placement.
- The usability of the ADAPT system had not been previously evaluated.
- This study aimed to assess the perceived usability of the ADAPT system in clinical practice.
Purpose of the Study:
- To investigate the perceived usability and user acceptability of the Stryker ADAPT computer-assisted navigation system.
- To evaluate the system's performance in assisting with Gamma3 nail placement for intertrochanteric femur fractures.
Main Methods:
- A descriptive study utilizing prospectively collected data from Aarhus University Hospital.
- Surgeons and nurses received training; ADAPT became mandatory for Gamma3 nail procedures.
- The System Usability Scale (SUS) questionnaire was administered post-procedure to assess perceived usability.
Main Results:
- Fifty procedures were performed by 29 surgeons using ADAPT, with a median first-time SUS score of 43, indicating "poor" usability.
- Experienced surgeons (≥3 procedures) showed no significant improvement in SUS scores over time.
- Positive feedback mentioned K-wire placement assistance and educational value; negative feedback cited inconsistency, slowness, and increased fluoroscopy time.
Conclusions:
- The majority of surgeons rated the usability and acceptability of the ADAPT system as "poor" or "not acceptable".
- Based on these findings, the ADAPT system has not been continued at the institution.
- The System Usability Scale is a viable tool for future research on navigation system usability in orthopaedic surgery.