Mechano-Node-Pore Sensing: A Rapid, Label-Free Platform for Multi-Parameter Single-Cell Viscoelastic Measurements
Published on: December 2, 2022
[Design of a dynamic measurement system for Bl factor spatial distribution of resonant blood viscoelastic sensor]
Jun Qian1,2, Haixuan Sun2,3, Bidou Wang2
1Division of Life Sciences and Medicine, School of Biomedical Engineering (Suzhou), University of Science and Technology of China, Suzhou, Jiangsu 215163, P. R. China.
Abstract:
Bl factor is a key system parameter of the resonant blood viscoelastic sensor. In this paper, a dynamic measurement system for the spatial distribution of Bl factor based on velocity amplitude and motional impedance was designed. The system extracted the velocity amplitude and motional impedance of the coil under the dynamic condition of driving the sensor to generate simple harmonic oscillations using laser displacement and impedance analysis combined with in-phase/quadrature demodulation algorithm, and controlled the equilibrium position of the coil by adjusting the direct current component of the excitation current to realize the position scanning. In the position interval of [-240, 240] μm, the maximum coefficient of variation of the measurement results was 0.077 3%, and the maximum relative error to the simulation results was 2.937 9%, with a linear fitting correlation coefficient R 2 = 0.996 8. The system can be used to accurately measure the spatial distribution of Bl factor of the resonant blood viscoelastic sensor, which provides a technical support for the verification of the design of the sensor magnetic circuit.
Related Concept Videos
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...