T-Wave Oversensing with Contemporary Implantable Cardioverter-Defibrillators
Marc Strik1,2, Sylvain Ploux1,2, Romain Eschalier3
1Bordeaux University Hospital (CHU), Cardio-Thoracic Unit, F-33600 Bordeaux, France.
Background:
Implantable cardioverter-defibrillators (ICDs) need to reliably detect ventricular tachycardia (VT) and ventricular fibrillation (VF) while avoiding T-wave oversensing (TWOS), which is associated with a risk of inappropriate therapies. The incidence of TWOS with endovascular ICDs appears to differ between manufacturers.
Objectives:
We aimed to evaluate the incidence and clinical consequences of TWOS with contemporary Medtronic and Boston Scientific ICDs.
Methods:
Consecutive patients implanted with a recent Medtronic or Boston Scientific ICD and remotely monitored at three French centers were included. All transmitted EGMs labelled as VF, VT, non-sustained VT (NSVT), or ventricular oversensing (Medtronic) were screened for TWOS.
Results:
Among 7589 transmitted episodes from 674 patients with a Boston Scientific ICD, we did not identify a single case of TWOS. Among 16,790 transmitted episodes from 1733 patients with a Medtronic ICD, we identified 60 patients (3.4%) with at least one episode of TWOS. In 46 patients, TWOS was intermittent (NSVT episodes). In the remaining 14 patients, TWOS resulted in 60 sustained episodes (completed counters). No inappropriate therapies were delivered in 12 of these patients because no therapies were programmed (in monitor zones, 11 episodes) or because therapies were inhibited by the morphology discriminator (Wavelet, 19 episodes) or by the anti-TWOS algorithm (26 episodes). Two patients received inappropriate therapies due to TWOS (0.1% of patients with Medtronic ICDs).
Conclusion:
On review of 24,379 transmitted episodes from 2407 patients with endovascular ICDs, we found no case of TWOS with Boston Scientific devices, whereas TWOS was not uncommon with Medtronic devices. However, the risk of inappropriate therapy with Medtronic ICDs was very low (0.1%) due to the often intermittent nature of this phenomenon, the morphology discriminator, and the anti-TWOS algorithm.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias VI: Management of Dysrhythmias
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
Dysrhythmias II: Classification of Tachyarrhythmias