Study on Mechanical Properties and Degradation Behavior of Magnesium Alloy Vascular Clip
Hongxu Zhang1,2, Ming Gao2, Xiaoying Tian2
1Materials Science and Engineering, Shenyang University of Chemical Technology, 11# Street Shenyang Economic & Technology Development Area, Shenyang 110142, China.
Abstract:
The Mg alloy vascular clip has biodegradability and good biocompatibility, which can improve the convenience and safety of clinical application. However, the Mg alloy vascular clip also has some disadvantages, such as an unreasonable structure design and a degradation rate which is too fast. In this study, the process of clamping blood vessels with a biodegradable Mg alloy (Mg-Zn-Nd-Zr and Mg-Zn-Nd) general V-type vascular clip was simulated by finite element simulation software (Abaqus). A new type of vascular clip, the P-type vascular clip, was analyzed and investigated through simulation. The differences between Mg alloy vascular clips of V-type and P-type were analyzed by finite element simulation. In addition, the effects of Zr element on the mechanical properties and corrosion resistance of P-type vascular clips were also investigated to improve the mechanical stability. The results show that during the V-type vascular clip closure of Mg-Zn-Nd-Zr alloy, this clip has some problems, such as uneven distribution of blood vessel stress, crevices in blood vessels and stress concentration. The improved P-type vascular clip has uniform closure, and there is no gap in the blood vessel, which can effectively avoid stress concentration. The improved P-type vascular clip is well closed and can effectively avoid stress concentration. The corrosion resistance of the Mg-Zn-Nd-Zr alloy P-type clip was better than that of the Mg-Zn-Nd alloy P-type clip (degradation rate of 2.02 mm/y and 2.61 mm/y on the 7th day, respectively). Mg-Zn-Nd-Zr alloy The P-type vascular clip remained closed even on the 7th day, which could meet the requirements of clinical application.
More Related Videos
12:18Co-localizing Kelvin Probe Force Microscopy with Other Microscopies and Spectroscopies: Selected Applications in Corrosion Characterization of Alloys
Published on: June 27, 2022
11:01Ferromagnetic Bare Metal Stent for Endothelial Cell Capture and Retention
Published on: September 18, 2015
Related Concept Videos
Mechanical Characteristics of Steel
The tension test is fundamental for determining tensile strength. In this test, a steel specimen is stretched using a gripping device until it breaks. The data collected during this test are used...
Fatigue