A Multicenter Comparison of 1-yr Functional Outcomes and Programming Differences Between the Advanced Bionics
Susan T Eitutis, Deborah A Vickers1, Karen Tebbutt2
1Sound Laboratory, Cambridge Hearing Group, Clinical Neurosciences, University of Cambridge, Cambridge.
Objective:
To determine if there is a difference in hearing outcomes or stimulation levels between Advanced Bionics straight and precurved arrays.
Study Design:
Retrospective chart review across three implant centers.
Setting:
Tertiary centers for cochlear and auditory brainstem implantation.
Patients:
One hundred fifteen pediatric and 205 adult cochlear implants (CIs) were reviewed. All patients were implanted under the National Institute for Health and Care Excellence 2009 guidelines with a HiRes Ultra SlimJ or Mid-Scala electrode array.
Main Outcome Measures:
Hearing preservation after implantation, as well as CI-only listening scores for Bamford-Kowal-Bench sentences were compared 1 year after implantation. Stimulation levels for threshold and comfort levels were also compared 1 year after implantation.
Results:
Hearing preservation was significantly better with the SlimJ compared with the Mid-Scala electrode array. Bamford-Kowal-Bench outcomes were not significantly different between the two arrays in any listening condition. Stimulation levels were not different between arrays but did vary across electrode contacts. At least one electrode was deactivated in 33% of implants but was more common for the SlimJ device.
Conclusion:
Modern straight and precurved arrays from Advanced Bionics did not differ in hearing performance or current requirements. Although hearing preservation was possible with both devices, the SlimJ array would still be the preferred electrode in cases where hearing preservation was a priority. Unfortunately, the SlimJ device was also prone to poor sound perception on basal electrodes. Further investigation is needed to determine if deactivated electrodes are associated with electrode position/migration, and if programming changes are needed to optimize the use of these high-frequency channels.
More Related Videos
Related Concept Videos
Comparing Mitochondrial, Chloroplast, and Prokaryotic Genomes
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
DNA Microarrays
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Mass Analyzers: Overview