The Influence of Laboratory Scanner versus Intra-Oral Scanner on Determining Axes and Distances between Three
Asaf Shely1, Diva Lugassy2, Ophir Rosner1
1Department of Oral Rehabilitation, The Maurice and Gabriela Goldschleger School of Dental Medicine, Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv 6997801, Israel.
This study compared laboratory scanners (LBS) and intra-oral scanners (IOS) for measuring dental implant accuracy. Scan abutment geometry influenced accuracy, with MIS ISB showing lower inter-implant distance changes and ZZ ISB showing lower intra-implant angle changes.
Area of Science:
- Digital Dentistry
- Dental Implantology
- 3D Scanning Technologies
Background:
- Accurate digital impressions are crucial for dental implant restorations.
- Comparing laboratory scanners (LBS) and intra-oral scanners (IOS) is essential for understanding their clinical applicability.
- Intra-oral scan bodies (ISBs) are key components in capturing precise implant data.
Purpose of the Study:
- To compare the accuracy of LBS versus IOS in measuring inter-implant distance, inter-implant axis, and intra-implant axis.
- To evaluate the influence of two different intra-oral scan bodies (MIS and Zirkonzhan) on scanning accuracy.
Main Methods:
- A 3D printed model with three implant analogs was used for in vitro scanning.
- Two types of titanium intra-oral scan bodies (MIS ISB and Zirkonzhan ISB) were scanned with LBS and IOS (Primescan).
- Stereolithography (STL) files were compared using 3D analysis software, with statistical analysis (Kolmogorov-Smirnov and Mann-Whitney tests).
Main Results:
- Both scanner types and ISBs showed variations in all measured parameters.
- The MIS ISB demonstrated significantly lower changes in inter-implant distance compared to the Zirkonzhan ISB.
- The Zirkonzhan ISB showed significantly lower changes in intra-implant angle compared to the MIS ISB.
Conclusions:
- The geometry of the intra-oral scan body significantly impacts the accuracy of digital implant measurements.
- Differences exist between LBS and IOS, and the choice of ISB affects specific dimensional and angular measurements.
- Further research is needed to explore the influence of scan abutment geometry, material, and components on overall scanning trueness.
Related Concept Videos
Methods for Studying Drug Absorption: In vitro
The diffusion cell method uses a two-compartment cell, including a donor compartment with the drug solution, which simulates the environment where the drug is applied, and a receptor compartment with a buffer solution, which simulates the environment...
Methods for Studying Drug Absorption: In situ
The Doluisio method involves perfusing a prepared segment of a rat's small intestine with a solution of radiolabeled drug and a non-absorbable marker. This helps to differentiate between absorbed and non-absorbed drug concentrations. The intestinal segment is connected at both ends using tubing and syringes,...