Repeatability and Agreement of Chord Mu between Scheimpflug Tomography and Swept-Source Optical Coherence Tomography
Na Hyun Kim1, Hyun Jin Kim1, Soo Chang Cho1
1Department of Ophthalmology, Ewha Womans University Mokdong Hospital, Ewha Womans University College of Medicine, Seoul, Korea.
Purpose:
To evaluate repeatability and agreement of chord mu between Scheimpflug tomography (Pentacam HR) and sweptsource optical coherence tomography-based optical biometer (IOLMaster 700).
Methods:
In this retrospective study, 63 eyes from 33 patients were included. Chord mu, X and Y Cartesian distances between the corneal vertex and the pupil center (Px and Py), and the pupil diameter were compared using two instruments. Repeatability was evaluated using intraclass correlation coefficient (ICC), coefficient of variation (CoV), and within-subject standard deviation (Sw). Interdevice agreement was evaluated using paired t-tests and Bland-Altman plots.
Results:
Although Sw values for all parameters were similar between the two devices, CoV values of chord mu and pupil diameter were lower, and ICC values of those parameters were higher, in the IOLMaster 700 than in the Pentacam HR. Chord mu and pupil diameter values were higher in IOLMaster 700 than Pentacam HR (p < 0.01). The width of the 95% limit of agreement was wide for all parameters.
Conclusions:
IOLMaster 700 showed better repeatability than Pentacam HR in chord mu, Px, Py, and pupil diameter values. Because there were statistically significant differences and a low level of agreement in chord mu and pupil diameter values between the two devices, they cannot be used interchangeably.
More Related Videos
12:54Simultaneous Brightfield, Fluorescence, and Optical Coherence Tomographic Imaging of Contracting Cardiac Trabeculae Ex Vivo
Published on: October 2, 2021
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography