Helmet continuous positive airway pressure for patients' transport using a single oxygen cylinder: A bench study
N Capsoni1,2, F Zadek2, D Privitera1
1Department of Emergency Medicine, ASST Grande Ospedale Metropolitano Niguarda, Milan, Italy.
Background:
Continuous positive airway pressure (CPAP) is frequently used to treat patients with acute respiratory failure in out-of-hospital settings. Compared to a facemask, the helmet has many advantages for the patient but requires a minimum gas flow of 60 L/min to avoid CO2 rebreathing. The aim of the present bench study was to evaluate the performance of four Venturi devices, connected to a single oxygen cylinder, in delivering helmet-CPAP with clinically relevant gas flow, fraction of inspired oxygen (FiO2), and positive end-expiratory pressure (PEEP) values.
Methods:
Three double-inlet Venturi systems (EasyVent, Ventuplus, Compact-HAR) were connected to full 5-L oxygen cylinders using a double flowmeter, and their oxygen requirements to reach different setups (flow 60-80 L/min; FiO2 0.4-0.5-0.6, PEEP 7.5-10-12.5 cmH2O) were tested. The fourth Venturi system (O2-MAX) was directly attached to the tank, and the flow and FiO2 delivered at preset FiO2 0.3 and 0.6 were recorded. The runtime of the cylinder was assessed.
Results:
EasyVent, Ventuplus, and O2-MAX were able to deliver helmet-CPAP with clinically useful setups when connected to a single oxygen cylinder, while Compact-HAR did not. The runtime of the cylinders ranged between 28 and 60 minutes according to the preset flow and FiO2. The delivered gas flow decreased slowly and linearly with the drop in cylinder pressure until its exhaustion.
Conclusions:
Helmet-CPAP might be provided using portable Venturi systems connected to an oxygen cylinder, but not all of them are able to deliver it. The use of a double flowmeter allows delivery of both high flow and high FiO2 when double-inlet Venturi systems are used. Due to the flow drop observed during the cylinder consumption, a flow >60 L/min should be set when helmet-CPAP is started. Considering the flow drop phenomenon, the estimated duration of the tank runtime can be used with a margin of safety when planning patient transport.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...