Analysis of the Ambient Particulate Matter-induced Chromosomal Aberrations Using an In Vitro System
Published on: December 21, 2016
Automatic weighing system vs. manual weighing precision comparison in PM-loaded filter measurements under different
Dmytro Chyzhykov1, Kamila Widziewicz-Rzońca2, Monika Błaszczak1
1Institute of Environmental Engineering, Polish Academy of Sciences, 34 M. Skłodowskiej-Curie St, 41-819, Zabrze, Poland.
Abstract:
Currently, there is a lack of research directly comparing the precision of automatic weighing systems and manual weighing in the context of particulate matter (PM) filter equilibration and measurements under different humidity conditions. During experimental measurements, three different types of PM-loaded filters were weighed using manual and automatic balances. During manual weighing, every filter was weighed twice in three different relative humidity conditions. The same procedure was done using an automated weighing system. In most cases, it was found that under relative humidities in the range of 30-55% RH, the manual and automated methods can be treated as referential. Regarding device stability, very slight but overall better precision was found for 30% RH, suggesting that 40 CFR Part 50, Appendix L requirements regarding conditioning humidity (30-40% RH) seem more suitable than those presented in the PN-EN 12341:2014 standard (45-50% RH). Understanding the effects of the influence of the RH% on PM mass measurements is a matter of great importance, because water vapor condensed on a filter can affect the particulate matter concentrations. This is especially important in areas where regulatory limits are exceeded. Calculation of uncertainty in the PM mass measurements is therefore crucial for determining the actual sample mass and improving air monitoring practices. In a nutshell, the experimental results obtained clearly describe how changing RH% conditions affect the PM weighing precision during manual and automated measurements.
More Related Videos
09:33Visualizing Field Data Collection Procedures of Exposure and Biomarker Assessments for the Household Air Pollution Intervention Network Trial in India
Published on: December 23, 2022
06:39A High Performance Impedance-based Platform for Evaporation Rate Detection
Published on: October 17, 2016
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Uncertainty in Measurement: Reading Instruments
Filtration
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...