Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
Noninvasive Point of Care Device for Assessing Cardiac Response to Acute Volume Changes
Harry A Silber1, Nisha A Gilotra1, Thomas L Miller2,3
1Department of Medicine, Division of Cardiology, Johns Hopkins University School of Medicine, Baltimore, MD, USA.
Purpose:
The change in the amplitude of a peripheral pulse in response to a Valsalva maneuver has diagnostic utility for assessing volume status at the bedside. We have developed a device to automatically quantify the Valsalva pulse response (VPR) to a standardized Valsalva maneuver that the device guides a user to perform. In this study, we sought to determine whether VPR by the device, Indicor, is sensitive enough to detect the acute increase in central pressure and volume load that occurs with a passive leg raise (PLR) in healthy volunteers.
Methods:
Healthy volunteers were tested semirecumbently at 45 degrees, then again after being leaned back on a pivoted wedge with legs raised at 45 degrees and torso and head flat, and then again in the semirecumbent position. The device recorded a finger photoplethysmography (PPG) signal during a 10-second expiratory effort of 20 mmHg as guided by the device. VPR was automatically calculated as the ratio of the end-Valsalva pulse amplitude to the baseline pulse amplitude.
Results:
In the 30 participants who completed testing, VPR increased from baseline to PLR in every participant, from 0.34 ± 0.13 to 0.60 ± 0.14 (p < 0.0001). Back upright, VPR decreased back to 0.33 ± 0.10 (p < 0.0001 versus PLR; NS versus baseline position).
Conclusion:
In this proof-of-concept study of healthy participants, the Indicor device, a noninvasive, convenient device that automatically calculates VPR from a finger photoplethysmography signal during a standardized Valsalva maneuver, was sensitive enough to detect the increase in VPR that occurred with an acute central volume load from a PLR. Future studies should examine whether VPR responds differently to a PLR in heart failure patients with abnormal cardiac performance and/or congestion.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: