Related Experiment Video
Updated: Jul 11, 2025

Novel Mini-open Transforaminal Lumbar Interbody Fusion
Published on: June 6, 2025
Percutaneous removal and replacement of a novel percutaneous interspinous device
Stefano Marcia1, Joshua Adam Hirsch2, Matteo Bellini3
1UOC RAdiologia, Ospedale SS Trinità, Cagliari, Italy.
Abstract:
Objective: The aim of this study is to explain the technique used for removing and replacing a novel percutaneous interspinous device (PID).Procedure: Three male patients, with a mean age of 66 years (range 62-72), were included in the study due to displacement (N = 2) or misplacement (N = 1) of the novel PID. This occurred after a mean of 23.3 ± 10.5 days from the initial placement (range 13-34) at the L4-L5 level (N = 1) and L3-L4 level (N = 2). Following the removal of the PID, four new devices were implanted.Conclusion: The novel PID can be safely removed either immediately after its placement during the procedure or after a certain period of days or months from its implantation. The removal procedure can be performed percutaneously under imaging guidance, by closing the four retractable wings, using the same instrumentation as utilized during the initial implantation.
More Related Videos
03:24Author Spotlight: Development and Application of a Novel Suture Technique for Annular Fibrosus Repair in Percutaneous Transforaminal Endoscopic Discectomy
Published on: January 26, 2024
12:25C-arm-Free Simultaneous OLIF51 and Percutaneous Pedicle Screw Fixation in a Single Lateral Position
Published on: September 16, 2022