Automated Acoustic Dispensing for the Serial Dilution of Peptide Agonists in Potency Determination Assays
Published on: November 10, 2016
A high-throughput quality control method for assessing the serial dilution performance of dose-response plates with
Mary Ashley Rimmer1, Nathaniel R Twarog2, Yong Li1
1Analytical Technologies Center, Department of Chemical Biology and Therapeutics, St. Jude Children's Research Hospital, Memphis, TN, United States.
Abstract:
This study aimed to develop a streamlined method for evaluating the dilution ratio of drug dose-response plates created by automated liquid handlers in the early stages of drug discovery. The quantitative techniques commonly used for this purpose have restrictions due to their limited linear dynamic range and inaccuracies in assessing serial dilution performance. To address this challenge, we describe a method based on acoustic ejection mass spectrometry (AEMS). The method involves using standard compounds and an internal standard to evaluate each dilution point in quality control (QC) plates. The samples are transferred to a chromatography-free tandem mass spectrometry system through an acoustic source, enabling the analysis of one sample per three seconds from a microtiter plate. This approach provides precise, accurate, label-free, and rapid data acquisition to support high-throughput screening efforts.