Preparing an Isotopically Pure 229Th Ion Beam for Studies of 229mTh
Published on: May 3, 2019
A Conjoined Rectilinear Collision Cell and Pulsed Extraction Ion Trap with Auxiliary DC Electrodes
Hamish Stewart1, Dmitry Grinfeld1, Alexander Wagner1
1Thermo Fisher Scientific, 11 Hannah-Kunath Str., 28199 Bremen, Germany.
Abstract:
Ion traps are routinely directly coupled to mass analyzers, where they serve to suitably cool and shape an ion population prior to pulsed extraction into the analyzer proper. Such devices benefit from high duty cycle and transmission but suffer slow ion processing times caused by a compromise in the buffer gas pressure range that suitably dampens the ion kinetic energy without causing excessive scatter during extraction or within the analyzer. A rectilinear RF quadrupole ion trap has been characterized, conjoining a pressurized collision region with a pumped extraction region, and an unbroken RF interface for seamless ion transfer between them. Auxiliary electrodes mounted between the RF electrodes provide DC voltage gradients that serve to both guide ions through the device and position them at the extraction slot. The influence of the auxiliary DC upon the trapping RF field was measured, and suitable parameters were defined. A mode of operation was developed that allowed parallel processing of ions in both regions, enabling a repetition rate of 200 Hz when the device was coupled to a high-resolution accurate-mass analyzer.
Related Concept Videos
Mass Analyzers: Common Types
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Ion-Exchange Chromatography
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...