Phantom study investigating the repeatability of radiomic features with alteration of image acquisition parameters in

Rosalind Mitchell-Hay1, Trevor Ahearn2, Alison Murray3

  • 1Aberdeen Biomedical Imaging Centre, University of Aberdeen, Aberdeen, Scotland; Radiology Department, NHS Grampian, Aberdeen, Scotland.

Abstract

Insights

Repeatability of radiomic features in MRI depends on imaging parameters. Careful selection of echo (TE) and repetition times (TR) is crucial for reliable radiomics analysis.

Area of Science:

  • Medical Imaging
  • Radiomics
  • Quantitative Imaging

Background:

  • Magnetic Resonance Imaging (MRI) parameters influence image appearance.
  • Radiomics extracts quantitative features from medical images.
  • Repeatability is a key limitation in radiomics.

Purpose of the Study:

  • Investigate the repeatability of radiomic features (RaF) in MRI.
  • Assess the impact of varying echo times (TE) and repetition times (TR) on RaF repeatability.

Main Methods:

  • A phantom was scanned 10 times under identical conditions on a 3T MRI scanner.
  • Scans were performed across a range of echo times (80-110 ms) and repetition times (3000-5000 ms).
  • Radiomic analysis was conducted on segmented phantom sections, with Intraclass Correlation Coefficient (ICC) used to assess repeatability.

Main Results:

  • 187 features showed high repeatability (ICC >0.9) across echo time variations.
  • 82 features demonstrated high repeatability (ICC >0.9) across repetition time variations.
  • Significant overlap (51 features) with ICC >0.9 was observed between TE and TR ranges.

Conclusions:

  • The repeatability of radiomic features in MRI is significantly influenced by imaging parameters.
  • Careful consideration of echo and repetition times is essential for robust radiomics analysis.
  • Variable selection strategies must account for parameter dependencies in MRI radiomics.