The quantitative aspect of photoplethysmography revised
Heart and Vessels
|January 1, 1986
Summary
Photoplethysmography (PPG) output is a logarithmic function of red cell density and intravascular volume. This study suggests a monologarithmic function describes PPG output, accounting for stasis effects on volume and hematocrit.
Area of Science:
- Biomedical Engineering
- Physiological Measurement
Background:
- Photoplethysmography (PPG) is widely used to measure physiological parameters.
- Understanding the transfer function of PPG is crucial for accurate data interpretation.
Purpose of the Study:
- To investigate the overall transfer function of PPG devices.
- To assess PPG output in relation to red cell density and intravascular volume.
Main Methods:
- In vitro assessment of PPG as a red cell densitometer.
- In vivo experiments correlating PPG output with air plethysmography measurements.
- Analysis of PPG output as a function of red cell density and intravascular volume.
Main Results:
- PPG output demonstrated a logarithmic relationship with red cell density, consistent with the Lambert-Beer theorem.
- In vivo, PPG output was found to be a logarithmic function of digital intravascular volume.
- A monologarithmic function was proposed to describe the combined effects of stasis on intravascular volume and hematocrit.
Conclusions:
- The study elucidates the non-linear transfer function of PPG devices.
- PPG output can be described by logarithmic and monologarithmic functions under different physiological conditions.
- Linearization of PPG electronics using an exponential amplifier is suggested for improved accuracy.