Related Experiment Video
Updated: Jul 11, 2025

Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Test method for vapor collection and ion mobility detection of explosives with low vapor pressure
Chae Eun Son1, He-Ryun Choi1, Sung-Seen Choi1
1Department of Chemistry, Sejong University, Seoul, Republic of Korea.
Rationale:
Ion mobility spectrometry (IMS) has been widely used for on-site detection of explosives. Air sampling method is applicable only when the concentration of explosive vapor is considerably high in the air, but vapor pressures of common explosives such as 2,4,6-trinitrotoluene (TNT), 1,3,5-trinitro-1,3,5-triazacyclohexane (RDX), and pentaerythritol tetranitrate (PETN) are very low. A test method for analyzing the vapor detection efficiency of explosives with low vapor pressure via IMS was developed using artificial vapor and collection matrices.
Methods:
Artificial explosive vapor was produced by spraying an explosive solution in acetone. Fifteen collection matrices of various materials with woven or nonwoven structures were tested. Two arrangements of horizontal and vertical positions of the collection matrices were employed. Explosive vapor collected in the matrix was analyzed using IMS.
Results:
Only three collection matrices of stainless steel mesh (SSM), polytetrafluoroethylene sheet (PFS), and lens cleansing paper (LCP) showed the TNT and/or RDX ion peaks at an explosive vapor concentration of 49 ng/L. There was no collection matrix to detect PETN vapor at or lower than 49 ng/L. For the PFS, TNT and RDX were detected at a vapor concentration of 49 ng/L. For the LCP, TNT and RDX were detected at vapor concentrations of 14 and 49 ng/L, irrespectively.
Conclusions:
The difference in the explosive vapor detection efficiencies could be explained by the adsorption and desorption capabilities of the collection matrices. The proposed method can be used for evaluating the vapor detection efficiency of hazardous materials with low vapor pressure.
Related Concept Videos
Atomic Absorption Spectroscopy: Atomization Methods
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Gas Chromatography: Types of Detectors-II
Volatilization
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
High-Performance Liquid Chromatography: Types of Detectors