Related Experiment Video
Updated: Jul 11, 2025

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Zero-fluoro atrioventricular-nodal reentrant tachycardia ablation
Christina Soether1, Andreas A Boehmer1, Bianca C Dobre1
1Department of Cardiology, St. Josefs-Hospital GmbH, Beethovenstraße 20, 65189, Wiesbaden, Germany.
Zero-fluoroscopy radiofrequency catheter ablation (RFCA) for atrioventricular-nodal reentrant tachycardia (AVNRT) is feasible and safe. Three-dimensional electroanatomical mapping (3D-EAM) reduces radiation exposure without increasing procedure time or complications.
Area of Science:
- Electrophysiology
- Cardiovascular Medicine
- Medical Imaging
Background:
- Atrioventricular-nodal reentrant tachycardia (AVNRT) is a common supraventricular tachycardia, particularly in younger individuals.
- Radiofrequency catheter ablation (RFCA) is the standard treatment for AVNRT.
- Traditional RFCA relies on ionizing radiation (fluoroscopy) for catheter guidance, posing risks to patients and staff.
Purpose of the Study:
- To evaluate the feasibility and safety of performing zero-fluoroscopy RFCA for AVNRT.
- To assess the utility of the EnSite™ NavX™ three-dimensional (3D) electroanatomical mapping system (EAM) in radiation-free ablation procedures.
Main Methods:
- Retrospective analysis of 68 patients undergoing AVNRT-RFCA.
- Comparison between conventional fluoroscopy (convFluoro, n=30), 3D-EAM with zero fluoroscopy (zeroFluoro, n=20), and 3D-EAM with converted fluoroscopy (convertedFluoro, n=18).
- Analysis of procedure duration, fluoroscopy time and dose, and complication rates.
Main Results:
- Procedure duration was similar between convFluoro and zeroFluoro groups (74±24 min vs. 80±26 min).
- 3D-EAM significantly reduced fluoroscopy duration (7±6 min in convertedFluoro vs. 12±9 min in convFluoro).
- ZeroFluoro cases involved no radiation; all procedures showed high success rates with few minor complications and no major adverse events.
Conclusions:
- Zero-fluoroscopy RFCA for AVNRT is a safe and feasible alternative to conventional methods.
- Utilizing 3D-EAM enables significant radiation reduction for most patients.
- This approach does not prolong procedure time or increase complication risks.
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
Antiarrhythmic Drugs: Class IV Agents as Calcium Channel Blockers
Verapamil, a calcium channel blocker, inhibits calcium movement across myocardial cell membranes and vascular smooth muscle. This results in the dilation of coronary and...