Efficacy and User Experience of a Novel X-Ray Shield on Operator Radiation Exposure During Cardiac Catheterization: A
Cedric Davidsen1,2, Kristian Ytre-Hauge3, Andreas Tefre Samnøy4
1Department of Heart Disease, Haukeland University Hospital, Bergen, Norway (C.D., K.V., V.T.).
Background:
Radiation shielding is mandatory during cardiac catheterization, but there is a need to improve efficacy and ease of use.
Methods:
The aim of the study was to assess the shielding effect and user feedback for a novel flexible multiconfiguration x-ray shield (FMX). The 0.5-mm Pb equivalent FMX can be selectively configured to accommodate for variations in patient morphology, access site, and type of procedure with maintained visualization, vascular access, and shielding. To evaluate efficacy, relative operator dose (operator dose indexed for given dose) was measured during 103 consecutive procedures randomized in a 1:1 proportion to the current routine setup or FMX+routine. User feedback was collected on function, relevance, and likelihood of adoption into clinical practice.
Results:
Median relative operator dose was 3.63 μSv/µGy·m2×10-3 (IQR, 2.62-6.37) with routine setup and 0.57 μSv/µGy·m2×10-3 (IQR, 0.27-1.06) with FMX+routine, which amounts to an 84.4% reduction (P<0.001). For 500 procedures/year, this corresponds to an estimated yearly dose reduction from 3.6 to 0.7 mSv. User feedback regarding size, functionality, ease of use, likely to use, critical issues, shielding, draping, procedure time, vascular access, patient discomfort, and risk was 99% positive. No critical issues were identified. There was no significant difference in patient radiation exposure.
Conclusions:
The FMX reduces radiation exposure considerably. The FMX represents an effective and attractive solution for radiation protection that can easily be implemented in existing workflow. FMX has potential for general use with maintained visualization, vascular access, and shielding in routine cardiac catheterization.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization III: Left Heart Catheterization
X-ray Imaging