An automatic pipeline for PET/MRI attenuation correction validation in the brain
Mahdjoub Hamdi1, Chunwei Ying2, Hongyu An2,3,4,5
1Mallinckrodt Institute of Radiology, Washington University in St. Louis, St. Louis, MO, USA. hamdi.m@wustl.edu.
EJNMMI Physics
|November 14, 2023
Summary
This study introduces an automated pipeline for evaluating MRI-based PET attenuation correction (AC) methods in neurological imaging. The DL-DIXON method demonstrated the least quantitative bias, offering improved accuracy for PET/MRI scans.
Area of Science:
- Neuroimaging
- Medical Physics
- Radiochemistry
Background:
- Accurate Positron Emission Tomography (PET) attenuation correction (AC) is crucial for quantitative accuracy in neurological PET/MRI applications.
- Existing MRI-derived PET AC methods face challenges in achieving precise quantitative results.
- The development of robust evaluation pipelines is necessary to assess novel AC techniques.
Purpose of the Study:
- To propose and evaluate an automated pipeline for assessing the quantitative accuracy of four MRI-derived PET AC methods.
- To utilize analytically simulated PET brain lesions and Regions of Interest (ROIs) as ground truth for PET activity.
- To compare the performance of DIXON, DIXONbone, UTE AC, and DL-DIXON against the gold standard PET CTAC.
Main Methods:
- An automated pipeline was developed, integrating a synthetic lesion insertion tool and the FreeSurfer neuroimaging framework.
- Simulated spherical lesions and ROIs were inserted into PET projection space and reconstructed using four MRI-based AC techniques.
- Quantitative accuracy was evaluated by comparing MRI-based AC to CTAC biases in lesions and ROIs, with and without background activity.
Main Results:
- The pipeline accurately assessed quantitative accuracy for simulated lesions and ROIs, consistent with original PET images.
- DL-DIXON exhibited the least bias (-0.08% to 0.22%) across evaluated conditions, followed by DIXONbone (-1.85% to -2.46%), UTE AC (-2.74% to -3.56%), and DIXON (-5.41% to -7.02%).
- The DL-DIXON method demonstrated superior quantitative accuracy compared to other MRI-derived AC techniques evaluated.
Conclusions:
- The proposed automated pipeline accurately evaluates MRI-derived PET AC methods without requiring measured PET emission data.
- The pipeline enables consistent generation of realistic lesion ROIs for assessing novel PET correction techniques.
- DL-DIXON shows the most promising results for accurate PET attenuation correction in neurological applications using MRI-derived data.