Elimination of spectral interference of 87Rb with 87Sr during ICP-MS measurements using a chromatographic technique
Tomohiro Narukawa1, Satoki Okabayashi2, Koichi Chiba2
1National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, 1-1-1 Umezono, Tsukuba, Ibaraki 305-8563, Japan. tomohiro-narukawa@aist.go.jp.
Abstract:
A new chromatographic separation technique was developed for the measurement of the strontium (Sr) isotope abundance ratio using inductively coupled plasma mass spectrometry (ICP-MS). This technique, LC-ICP-MS, utilizes the addition of 18-crown-the eluent and an ODS column. The technique completely eliminates the spectral interference of 87Rb (an isobar of 87Sr) with 87Sr during the ICP-MS measurement. 18-Crown-6 selectively forms a clathrate compound with some metals. When the Sr isotope abundances are measured using LC-ICP-MS with an ODS column, the addition of a specific amount of 18-crown-6 to the HNO3/0.1% methanol eluent elongates the retention time of Rb. Therefore, Rb can be completely separated from Sr using LC because Rb is captured into the 18-crown-6 molecule and forms a clathrate compound. The technique enables the measurement of Sr isotope abundance ratios easily, even in the presence of a large amount of Rb. The proposed technique was applied to the measurement of Sr isotope abundance ratios in rice samples. To validate the LC-ICP-MS technique, a fraction containing Sr was collected, and the 87Sr/86Sr isotope ratio was precisely measured using ICP-MS.
More Related Videos
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
High-Resolution Mass Spectrometry (HRMS)
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...