On-line sample preconcentration by LVSEP-FASI with simple voltage control on Y-channel chips
Fumihiko Kitagawa1, Takahiro Yonekura2, Isoshi Nukatsuka2
1Department of Frontier Materials Chemistry, Graduate School of Science and Technology, Hirosaki University, 3 Bunkyo-cho, Hirosaki, Aomori, 036-8561, Japan. kitagawa@hirosaki-u.ac.jp.
Abstract:
To eliminate complicated voltage controls for highly sensitive microchip electrophoresis (MCE) analyses on the basis of combining two online sample preconcentration techniques, large-volume sample stacking with an electroosmotic flow (EOF) pump (LVSEP) and field-amplified sample injection (FASI), cross-channel microchips and a multichannel high-voltage power supply were replaced to Y-channel chips and a conventional power supply designed for capillary electrophoresis, respectively. By simple switching of the electric circuit after the LVSEP-FASI sample enrichments, the focused analytes could be separated during anodic migration in a separation channel. In the LVSEP-FASI analysis of fluorescein using the Y-channel microchip, the maximum sensitivity enhancement factor (SEF) of 7400 was achieved, resulting in a 30-fold detectability increase compared to the conventional LVSEP. The developed method was applied to the oligosaccharide analysis in MCE. As a result, the SEF for maltotriose was improved from 450 to 2300 and the baseline separation of the oligosaccharides was achieved without any complicated voltage control in LVSEP-FASI on the Y-channel chips.