Related Experiment Video
Updated: Jul 11, 2025

Post Column Derivatization Using Reaction Flow High Performance Liquid Chromatography Columns
Published on: April 26, 2016
On-column modification for the creation of temperature-responsive stationary phases
Adriaan Ampe1, Elena Bandini1, Ken Broeckhoven2
1Separation Science Group, Department of Organic and Macromolecular Chemistry, Krijgslaan 281, Ghent, Belgium.
Abstract:
Temperature-responsive liquid chromatography (TRLC) offers an alternative for retention and selectivity optimisation in HPLC. This approach thereby exploits temperature gradients on stimuli-responsive stationary phases and forfeits the necessity for solvent gradients, allowing analyses to be performed using aqueous mobile phases. Consequently, it can be employed as a green alternative to reversed-phase separations. However, current production to obtain temperature-responsive columns inherently require dedicated column packing processes with polymer-modified particles. To facilitate the development of temperature-responsive phases, a flow-through modification procedure was developed allowing on-column modification of aminopropyl silica columns. Three columns were manufactured using this novel flow-through approach, which achieved identical column efficiencies compared to existing TRLC column. Demonstrating the possibility of bypassing the dedicated packing processes without losing efficiency. Additionally, it was observed that flow-through produced columns yielded higher retention at elevated temperatures despite their reduced carbon load. Further investigation of the carbon load revealed the presence of stationary phase gradients, whose influence was studied via novel developed retention experiments, which revealed a negligible change reduction in retention upon a change of polymer modification from 19.8% to 14.5%. However, further decrease from 14.5% to 12.3% resulted in a larger change. Interestingly, a further enhancement in apparent plate numbers was observed when operating the column under a reversed flow, yielding an increasing stationary phase gradient. This on-column modification procedure demonstrates the potential for modification of existing (commercial) packed columns to achieve temperature-responsive phases without loss of efficiency or retention. Thus, not only facilitating accessibility to temperature-responsive phases, but also aiding with development of further generations of temperature-responsive phases by removing the need for packing optimisation. Additionally, a novel experiment was set up to easily investigate the effect of inhomogeneous stationary phases retention.
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Principles Of Column Chromatography
Chromatography: Introduction
The phase in which the compounds linger or on which the compounds adsorb is called the stationary phase, whereas the mobile phase is the solvent that carries the solutes to be analyzed. In traditional column chromatography, the mixture flows through the stationary phase, and the compounds partition between the stationary and mobile phases...
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...