Influence of ionization volume and sample gas flow rate on separation power in gas chromatography-ion mobility
Tim Kobelt1, Martin Lippmann1, Jannik Wuttke1
1Department of Sensors and Measurement Technology, Institute of Electrical Engineering and Measurement Technology, Leibniz University Hannover, Appelstr. 9A, Hannover 30167, Germany.
Abstract:
In this work, the influence of the sample gas flow rate and the ionization region volume of an ion mobility spectrometer (IMS) used as a detector in gas chromatography (GC) on GC-IMS peak shape has been investigated. Therefore, a drift tube IMS with a field-switching ion shutter, a defined ionization region volume and an ultra-violet radiation source was used. To identify the influence of the sample gas flow rate entering the ionization region (equals the GC carrier gas flow rate if no further make-up gas is used) and the ionization region volume on peak broadening and signal intensity, different sample volumes as they would elute from a GC were tested at a variety of sample gas flow rates at a given ionization region volume. The results clearly show that for low sample gas flow rates a depletion of sample molecules in the ionization region leads to a significant decrease in effective detector volume but also to reduced signal intensities. Therefore, for optimal performance of a GC-IMS, the optimal operating point of the GC should match the flow range, where the IMS provides the best compromise between signal-to-noise ratio and peak broadening.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
11:18A Practical Guide on Coupling a Scanning Mobility Sizer and Inductively Coupled Plasma Mass Spectrometer SMPS-ICPMS
Published on: July 11, 2017
Related Concept Videos
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Analyte Adsorption and Distribution
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...