Robotic Cochlear Implantation for Direct Cochlear Access
Published on: June 16, 2022
Comparative study of vestibular function preservation in manual versus robotic-assisted cochlear implantation
Arthur Derieppe1, Adrien Gendre1,2, Kinnie Bourget-Aguilar1
1Service d'Oto-Rhino-Laryngologie et chirurgie cervico-faciale, CHU Nantes, Nantes, France.
Objective:
To compare vestibular outcomes in cochlear implant (CI) surgery, between robotic-assisted insertion of the electrodes versus manual insertion.
Methods:
We performed a monocentric retrospective study. From March 2021, the robotic system RobOtol© was used for all CI cases. We compared this robotic-assisted insertion group with a manual insertion group of patients who received a CI between July 2020 and March 2021. Primary objective was vestibular outcome. We used objective vestibular function tests: caloric testing, Vestibular Evoked Myogenic Potential (VEMP), and Video Head Impulse Test (VHIT). Secondary objectives were postoperative complications including patient-reported postoperative vertigo.
Results:
We found no statistically significant difference between the two groups in terms of caloric testing, VEMP or VHIT outcomes. In patient-reported outcomes, there was significantly more vertigo in the manual insertion group compared with robotic-assisted insertion.
Conclusion:
It is hypothesized that a non-traumatic insertion would cause less vestibular dysfunction postoperatively. Larger prospective studies are required to determine whether robotic-assisted CI insertion has a significant impact on vestibular outcomes in CI surgery.