Convex triangular vs. cylindrical field of view: how does the shape of the FOV affect radiation dose?
Deivi Cascante-Sequeira1, Christiano Oliveira-Santos2, Danieli Moura Brasil2
1Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba, Brazil. dantikes@gmail.com.
Objective:
To compare the dosimetry between convex triangular fields of view (FOV) and similar dimension cylindrical FOVs of two cone-beam computed tomography (CBCT) models.
Methods:
Optically stimulated luminescence dosimeters (OSLDs) were placed in fiducial anatomical locations in an anthropomorphic phantom representing an adult head male for dosimetry scans. Convex triangular FOVs (100 × 80 mm/maxilla-mandible; 100 × 50 mm mandible; 100 × 50 mm/maxilla) from Veraviewepocs 3D R100 (J. Morita, Kyoto, Japan) (R100) and Veraview® X800 (J. Morita, Kyoto, Japan) (X800) and cylindrical FOVs from R100 and X800 (80 × 80 mm/maxilla-mandible; 80 × 50 mm/mandible; 80 × 50 mm/maxilla) were obtained, resulting in 12 different scan protocols. Equivalent doses for each relevant organ/tissue and the effective dose for each protocol were calculated. Mean effective doses were compared by the two-way analysis of variance (ANOVA) with Tukey's post hoc test to evaluate the effect of the FOV and device (α = 0.05).
Results:
The effective doses ranged between 69 and 324 μSv for the convex triangular FOVs and 76 and 332 μSv for the cylindrical FOVs. Convex triangular FOVs from the R100 device had effective doses 2.3 to 15.3% lower than their corresponding cylindrical FOVs with similar height (p < 0.05), and that difference ranged between 8.8 and 11.8% for the X800 device (p < 0.05).
Conclusion:
Convex triangular fields of view delivered slightly lower effective doses than the cylindrical fields of view of similar dimensions in the R100 and X800 CBCT devices.
Clinical Relevance:
Understanding the influence of the image geometry formation in effective dose allows optimization to reduce patient dose.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Biological Effects of Radiation
Spherical and Cylindrical Capacitor
Conventionally, considering the symmetry, the electric field between the concentric shells of a spherical capacitor is directed radially outward. The magnitude of the field,...
Gauss's Law: Cylindrical Symmetry
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Deformation in a Circular Shaft
Electric Field of a Non Uniformly Charged Sphere
Consider a non-uniformly charged sphere, for which the density of charge depends only on the distance from a point in space and not on the direction. Such a sphere has a spherically symmetrical charge distribution. Here, the electric...