Detection of blasts using flags and cell population data rules on Beckman Coulter DxH 900 hematology analyzer in
Hanah Kim1, Mina Hur1, Jong-Ho Yi1
1Department of Laboratory Medicine, Konkuk University School of Medicine, Seoul, Korea.
Objectives:
White blood cell (WBC)-related flags are essential for detecting abnormal cells including blasts in automated hematology analyzers (AHAs). Cell population data (CPD) may characterize each WBC population, and customized CPD rules can be also useful for detecting blasts. We evaluated the performance of WBC-related flags, customized CPD rules, and their combination for detecting blasts on the Beckman Coulter DxH 900 AHA (DxH 900, Beckman Coulter, Miami, Florida, USA).
Methods:
In a total of 239 samples from patients with hematologic diseases, complete blood count on DxH 900 and manual slide review (MSR) were conducted. The sensitivity, specificity, and efficiency of the five WBC-related flags, nine customized CPD rules, and their combination were evaluated for detecting blasts, in comparison with MSR.
Results:
Blasts were detected by MSR in 40 out of 239 (16.7 %) samples. The combination of flags and CPD rules showed the highest sensitivity compared with each of flags and CPD rules for detecting blasts (97.5 vs. 72.5 % vs. 92.5 %). Compared with any flag, the combination of flags and CPD rules significantly reduced false-negative samples from 11 to one for detecting blasts (27.5 vs. 2.5 %, p=0.002).
Conclusions:
This is the first study that evaluated the performance of both flags and CPD rules on DxH 900. The customized CPD rules as well as the combination of flags and CPD rules outperformed WBC-related flags for detecting blasts on DxH 900. The customized CPD rules can play a complementary role for improving the capability of blast detection on DxH 900.