Related Experiment Video
Updated: Jul 10, 2025

Ultrasound Velocity Measurement in a Liquid Metal Electrode
Published on: August 5, 2015
Simultaneous Measurement of Flow Velocity and Electrical Conductivity of a Liquid Metal Using an Eddy Current Flow
1Helmholtz-Zentrum Dresden-Rossendorf, Bautzner Landstraße 400, 01328 Dresden, Germany.
Abstract:
The Eddy Current Flow Meter (ECFM) is a commonly employed inductive sensor for assessing the local flow rate or flow velocity of liquid metals with temperatures up to 700 ∘C. One limitation of the ECFM lies in its dependency on the magnetic Reynolds number for measured voltage signals. These signals are influenced not only by the flow velocity but also by the electrical conductivity of the liquid metal. In scenarios where temperature fluctuations are significant, leading to corresponding variations in electrical conductivity, it becomes imperative to calibrate the ECFM while concurrently monitoring temperature to discern the respective impacts of flow velocity and electrical conductivity on the acquired signals. This paper introduces a novel approach that enables the concurrent measurement of electrical conductivity and flow velocity, even in the absence of precise knowledge of the liquid metal's conductivity or temperature. This method employs a Look-Up-Table methodology. The feasibility of this measurement technique is substantiated through numerical simulations and further validated through experiments conducted on the liquid metal alloy GaInSn at room temperature.
More Related Videos
Related Concept Videos
Pipe Flowrate Measurement
The orifice meter is a simple,...
Eddy Currents
Other major applications of eddy currents appear in metal detectors and the braking systems of trains and roller...
Theory of Metallic Conduction
In this theory, Newton's second law of motion is used to determine the acceleration of an electron in the presence of an applied electric field. Then, its velocity is expressed via this acceleration.
An electron moves through the crystal, containing positive ions,...
Controlled-Current Coulometry: Overview
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
Flow Table Test
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...