Techniques and Limitations of Gore Thoracoabdominal Multibranch Endoprosthesis (TAMBE) for Type 1A Endoleak After
Sukgu M Han1, Alexander D DiBartolomeo1, Marin Chavez2
1Division of Vascular Surgery and Endovascular Therapy, Keck Medical Center of University of Southern California, Los Angeles, CA, USA.
Purpose:
Endovascular aortic aneurysm repair (EVAR) is the dominant treatment modality over open repair for abdominal aortic aneurysms. However, a higher rate of reinterventions remains the Achilles heel of EVAR. Although type 1A endoleak from proximal seal zone failure of EVAR remains one of the leading causes for reintervention, fenestrated branched devices suitable for proximal extension of failed EVAR are not widely available in the United States. Gore Thoracoabdominal Multibranch Endoprosthesis (TAMBE) is an off-the-shelf investigational device that provides supraceliac seal by incorporating 4 visceral and renal arteries via preloaded inner branches.
Case Report:
In this article, we describe 2 cases of type 1A endoleak from previous EVAR devices repaired using TAMBE. Both cases were performed under the Food and Drug Administration (FDA) compassionate use exemption. Considerations on the case planning and implantation techniques of TAMBE specific to previous EVAR devices are reviewed.
Conclusions:
Gore TAMBE can be utilized to repair a type 1A endoleak of a previous infrarenal EVAR device. Greater supraceliac coverage necessary for TAMBE relative to the minimal seal zone should be considered when applying this device for a type 1A endoleak.
Clinical Impact:
This report demonstrates the feasibility of applying off-the-shelf TAMBE device to treat one of the most common failure modes of EVAR, type1A endoleak.